Customization: | Available |
---|---|
Warranty: | 1 Year |
Element: | Metals Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Test Items | Measuring Range | Unit | |
pH(CH) | 6.000~9.000 | pH scale | |
pCO2 | 1.607~26.67 | kpa | |
8.0~200.0 | mmHg | ||
pO2 | kpa | kpa | |
0~800.0 | mmHg | ||
K | 0.5~15.0 | mmol/L | |
Na+ | 20.0~200.0 | mmol/L | |
Cl- | 20.0~200.0 | mmol/L | |
Ca++ | 0.3~5.0 | mmol/L | |
Hct | 12.0~65.0 | % | |
B.P. | 500~800 | ||
Glu | 1.1~66.7 | mmol/L | |
Lac | 0.4~30.0 | mmol/L | |
Derived(Calculated) Parameters | |||
pH(TC) | pH Temperature Corrected | ||
pCO2(TC) | pCO2 Temperature Corrected | ||
pO2(TC) | pCO2 Temperature Corrected | ||
TCO2 | Total Carbon Dioxide | ||
HCO3 | Actual Plasma Bicarbonate | ||
BEb | Base Excess in Blood | ||
BEecf | Base Excess Inextra Cellular Fluids | ||
SBC | Standard Bicarbonate | ||
So2% | Calculated Oxygen Staturation Percentage | ||
Ca++(7.4) | Ionized Calcium Normalized'' to pH 7.4 | ||
TCa | Total Calcium Content | ||
AG | Anion Gap | ||
RI | Respiratory Index | ||
AaDO2 | Alveolar Artillery Oxygen Gradient | ||
O2SAT | Oxygen Saturation | ||
tHb(c) | Hernoglobin (Calculated) | ||
Model | BGE-800 Series |
Input Parameters | Patient Temperature, Hemoglobin, FiO2, B.P. |
Analysis Time | <90s |
Sample Volume | 95μl (whole blood), 35μl (capillary blood) |
Calibration | Automatic, Programmable, One-point and Two-point |
Data Storage | More than 10,000 results |
Display | 10.4'' TFT Touch Screen |
Battery | Leadacidbattery, 2.3Ah |
Printer | Build inthermalprinter, 70mm |
Barcode Reader | Up to 14 Kindsbarcodetype |
Interface | Serialline RS-232 |
TCP/IPinterface | |
USBport | |
Integratedbar-code reader | |
Working Conditions | PowerSupply: AC220±10% 50/60Hz; 110±10% 60Hz;150W |
Temperature: 15ºC~30ºC | |
Storage Environment: 5~45ºC | |
Package Size(W*D*H) | 800*680*640mm |
Gross Weight | 50kg |
ReagentPackage Size | 320*460*370mm |
ReagentGross Weight | 7kg |